Operating the device
The system software of the gas chromatographs CGA21 is
based on Windows® . To a large extent
self-explanatory, its context sensitive help system allows the device to be
operated without using the written documentation a lot. However, basic
knowledge of the principles of gas chromatography is indispensible as well as
observance of the generally accepted safety rules, in particular if hydrogen is
used when operating the FID.
After turning on the device, different power on
messages may appear.
In
the heated state, TCD units can be damaged even by minute traces of oxygen.
For TCD units to provide trouble-free operation, you must therefore ensure that
the supply of carrier gas is sufficient at all times (refer to "Summary of the
essential properties of a TCD:")
If a TCD unit is turned on, it will issue a
corresponding warning message. It is only after you have acknowledged this
message with OK that the heating of the TCD is activated; in this manner, you
can ensure a sufficient supply of carrier gas without causing any damage to the
detector.
As a rule, the flame ionisation
detector is heated to its minimum temperature of 150°C immediately after the
device has been activated, to prevent water from condensing in the detector. At
this high temperature, it is no longer possible to measure gas flows; for this
reason, the message displayed on the right is issued immediately after the GC
has been turned on. If you then select „Measure”, the detector remains unheated
until the first measurement has been started. This allows you to dismantle the
unheated detector and measure the gas flows first. If you select OK instead,
the initial menu of the device is called up. Measuring gas flow is described in
chapter 6.

Main
window
Functional components of the main window:
· Menu bar with temperatures
· Main display
· Peaklist
· Status bar
The menu bar buttons trigger basic actions. Rollover with mouse cursor pop up hints.
|
|
New Measurement: see chapter „New measurement“ |
|
|
Open measurement: Opens existing measurement. Supports older binary *.gcd files as well as the current XML format *.gcx. |
|
|
Save measurement: Stores existing measurement in XML file format (file extension: .gcx) In general it is recommended to store measurements
on USB flash disk and not on the system disk. |
|
|
Comparison of chromatograms: Allows you to compare a chromatogram with one or
more reference chromatograms. For details see chapter 2.4 “Evaluation of a
chromatogram“. This button toggles
between two states. |
|
|
Print report: Print report as pdf–file.
The report includes the temperature program, user data, main display with
currently selected section and the peak list. Although possible, we do not
recommend using a printer with a CGA21. It is better practice to take
measurement files to a PC workstation on an USB flash drive and make
printouts there. As an alternative you can make pdf
files at the GC, save it to USB flash disk and make printouts at a PC
workstation. |
|
|
Information about the current measurement: Displays information about the open measurement,
such as temperature program, user data etc. User data can be edited. |
|
|
Save main display to file:.
Saves an image of the
main display into a emf –file. This allows for convenient insertion of
GC curves into every common office program. |
|
|
Help. Access to common help. Important windows
contain additional context sensitive help buttons. |
|
|
Measuring gas flows Utility for convenient measurement of gas flows using the soap bubble
meter. |
|
|
Options: Reserved for future use |
|
|
|
The display allows
for easy visualization and evaluation of GC measurements. Use the „Open
Measurement“ dialog in the usual manner. The
chromatogram will be scaled automatically to fit the display area.
Symbol colors:
|
Red: |
Regular GC curve |
|
Blue: |
Peaks |
|
Green: |
Peaks
with a known mass are colored green. They can be used to determine the mass
of peaks with unknown mass, provided it is the same compound. |
|
Yellow: |
Indicates
marked peaks. Marked peaks can be deleted using the „delete peak“ button. |
Using the main display (GC Graph):
Pan: simply move the
mouse while pressing down the right
mouse button. Alternatively use the arrow buttons.

Select zoom area

Enlarged
section
Zoom: Draw a rectangle
from top-left to bottom-right while holding down the left mouse button as shown above. Mouse gesture in the opposite
direction (bottom-down → top, left) restores standard view.
|
Actions: |
|
|
Create a peak: |
Peaks can be created by double clicking into the display. The
first double click creates a starting point which is indicated by an arrow
labeled „start“, the second double click marks the endpoint of a peak. The
second double click must be at higher time values than the first. |
|
Mark peak: |
Double click into the
(blue) peak: Peak will be marked and
its color changes to yellow. More than one peak can be marked |
|
Divide
Peak: |
Peaks can be divided.
Left click button „Divide peak“, the mouse cursor turns into a vertical line.
The next left click divides the peak under the line cursor. |
|
Delete peak: |
A left click onto button
„delete peak“ will erase all marked peaks. |

Peaklist
Marked
peaks will be added to the peaklist
immediately. The entries of the peaklist will be
transferred into the measurement file when a „Save“ command
is executed. Every line in the peak table represents one peak with its
corresponding properties. Fig. 5 shows a mixture of alkanes as an example:
|
The columns of the peak table: |
|
|
Nr.: |
Peak number in ascending order of their retention times. |
|
Name: |
Name of peak given by user. To assign a name to a peak, left click
into the peak. The table now shows this peak in the first line highlighted in
light yellow. After a double click into the name field a window pops up where
the name can be entered. |
|
Max[s]: |
Position of maximum: Elapsed time between injection and peak maximum. Time in seconds. |
|
Area: |
The Peak area is proportional to mass for a given substance. It is
determined automatically by numerical integration in arbitrary units. Please
note: The relationship between mass and area of a peak is not trivial, it depends among others from the kind of detector
used, the substance and the split ratio. As a consequence these values are
only useful for comparison. |
|
Area[%]: |
Percentage share of peak area in the sum of all marked peak areas. If
only one peak is marked, this percentage is 100%. |
|
Mass and factor: |
If the peak has been standardized with a reference measurement, this
column shows the mass in [ng] and the factor for
conversion of peak area into mass – and
vice versa. These columns can be used in two different ways: |
Displays
information about the state of the chromatograph: (from left to right)
·
Textual information about the current state of the
device
·
TCD Thermal Conductivity Detector on/off
·
Measurement
progress
·
Detector
type
·
Time elapsed since the GC has been turned on.