How to
conduct a GC measurement
Clicking „New measurement“ shows
the parameter window:

Fig. 6: Parameter window
The temperature program of the column furnace is set
in the upper section of the window. Basically, two different types of
measurement can be configured here:
1. The initial temperature at which the
measurement is starting.
2. The final temperature at the end of the
heating phase.
3. The start of the heating phase,
indicating the length of the preceding, isothermal phase.
4. The heating rate, i.e. the rate in
degrees/minute at which the temperature is to rise during the heating phase.
The individual values of the temperature program can
be set with the arrow keys next to each field. The entered values are checked
for plausibility, and dependent values are corrected
accordingly. For example, if the final temperature you have selected is
lower than the initial temperature, then the initial temperature is lowered
automatically to the same value as the final temperature
The minimum temperature of the detector is 150°C. At
higher final temperatures, the detector is heated accordingly before
commencement of the measurement. The temperature of the injector is 30°C higher
than the selected final temperature. You can change this property by unchecking
the “Determine Injector temperature automatically” checkbox. In this case, the
detector temperature has to be entered manually. Click on OK to accept the
displayed values, click on Cancel to abort preparations for the measurement.
The column furnace, detector and injector now need
time to reach the set temperatures. Several different messages can appear now,
depending on the instantaneous temperature of the furnace.
Input
of user data
The lower section of this window contains several text
fields where important information concerning the measurement can be entered.
However, this information is optional and does not affect the measurement
itself and can also be changed subsequently if required. The detector type is
set by the system and cannot be changed
Loading
and saving methods
By
pushing the “Save method” button the contents of the parameter window can be
stored into a method file (file extension .gcm) while
the “Load method” button allows for restoring these values from file. In both
cases the usual file dialog is used. It is advised to have a separate folder
containing method files. Instructions for sample preparations can be integrated
into the HTML – help system.
Closing the parameter window with the
„OK“ button initiates the start
procedure of the measurement.

Dialog prepare
measurement
Before the actual measurement can take place, it is
necessary for the column and the detector to reach their final temperatures. You
will be notified about this in the same window. You can also start the
measurement immediately, in this case, however,
measurement is made with an invalid temperature program.
After the dialog has changed to the following message,
all temperatures are valid.

Dialog ready
for injection
Now it is time to fill the microliter syringe with the
substance to be analysed and adjust the volume to the
required value. Once all the preparations for the measurement have been
completed, click on “start meas.”. In
the picture then a number appears displaying a countdown sequence which gives
you 10 seconds time to insert the needle of the syringe through the septum.
Once the countdown has elapsed to zero, inject the test substance. Then
withdraw the microliter syringe from the injector. If a high injector
temperature has been selected, it is advisable to insert the needle of the
syringe as late as possible - depending on how practised
you are - in order to prevent premature vaporisation
of the test substance from the needle. In this context, also refer to the
chapter 3.2 Notes on Injection
The actual measurement procedure can now be commenced.
The components of the test mixture drift at different rates through the
separation column. Their emergence from the column is registered electronically
and recorded at certain time intervals. At the same time, these values are
plotted as a characteristic on the screen. The display of this characteristic
can be adjusted during the measurement.
Without intervention by the operator, the device
automatically terminates the measurement after the scheduled time. However, you
can terminate the measurement manually at any time by clicking “Stop
measurement”.
In the case of TCD units, it is also possible to shift
the reference line with the buttons labelled “baseline
up” and “baseline down”, e.g. to correct drifts caused by the detector. This
function is not available for FID units, due to their negligible drift.

The functional principle of the FID is based on ionisation
of hydrocarbon fragments inside a hydrogen flame. Prior to any measurement the
FID therefore has to be ignited. As the hydrogen flame is operated with an
excess of combustion air, ignition is only successful if the excess of air is
reduced for a short time. Ignition can be visualized because the detector
current changes dramatically in the course of this process. Simply start a new
measurement before ignition and take a closer look at the resulting curve,
which is didactically interesting too:
Igniting
the Flame Ionisation Detector
·
Wait until the detector temperature is higher than 100
°C.
·
Memorize (or better write down!) the position of the
hydrogen valve.
·
Turn the valve ¼ turn counter clockwise.
·
Fan the flame of a gas lighter over the top of the
FID. Ignition is indicated by a small oxyhydrogen
explosion.
·
Don’t
forget to turn back the hydrogen valve to its former position.